Modulating Source
The function generator will accept an internal modulating signal,
an external modulating signal, or both.
Modulating source: Internal-External (both) or External only.
The default is Both (internal-external).
The External modulating source is always enabled.
When both sources are enabled (
internal-external
), the function
generator adds the internal and external modulating signals
(the carrier waveform is actually modulated with two waveforms).
When the internal source is disabled (
external only
), the carrier
waveform is modulated with the external waveform. The
Ext
annunciator turns on to indicate that the function generator is
expecting a modulating signal on the rear-panel
AM Modulation
terminal.
You apply the external modulating waveform to the
AM Modulation
terminal. The
modulation depth
is controlled by the signal level
present (5 volts
peak
corresponds to 100% modulation).
Front-Panel Operation
:
2: AM SOURCE
(MOD MENU)
Remote Interface Operation
:
AM:SOURce {BOTH|EXTernal}
5 V
0 V
-5 V
3
Chapter 3 Features and Functions
Amplitude Modulation (AM)
75
Содержание 33120A
Страница 2: ......
Страница 10: ...8 ...
Страница 14: ...Contents Contents 12 ...
Страница 15: ...1 Quick Start 1 ...
Страница 31: ...2 Front Panel Menu Operation 2 ...
Страница 55: ...3 Features and Functions 3 ...
Страница 126: ...124 ...
Страница 127: ...4 Remote Interface Reference 4 ...
Страница 204: ...SCPI Status System Chapter 4 Remote Interface Reference The SCPI Status Registers 202 ...
Страница 228: ...226 ...
Страница 229: ...5 Error Messages 5 ...
Страница 244: ...242 ...
Страница 245: ...6 Application Programs 6 ...
Страница 272: ...270 ...
Страница 273: ...7 Tutorial 7 ...
Страница 299: ...8 Specifications 8 ...
Страница 304: ...302 ...