Agilent 1260 Infinity DAD and MWD User Manual
91
Using the Detector
4
Special Settings of the Detector
Margin for Negative Absorbance Settings
1
To change the settings select
Setup Detector Signals
.
2
In the section Margin for Negative Absorbance change the value according
to your needs.
Optimizing the Detector
Additional theoretical information can be found in chapter
Use this field to modify the detector’s signal handling to increase
the margin for negative absorbance. Use this option if, for
example, your solvent gradient produces a decreasing baseline
absorbance, and for GPC analyses.
Limits
: 100 to 4000 mAU.
Figure 31
Margin for Negative Absorbance
N O T E
The higher the value the greater the baseline noise. Set this value only if you expect
negative absorbance greater than -100 mAU.
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......