![Agilent Technologies 1260 Infinity G1315C Скачать руководство пользователя страница 269](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-g1315c/1260-infinity-g1315c_user-manual_2867535269.webp)
Agilent 1260 Infinity DAD and MWD User Manual
269
LAN Configuration
13
Dynamic Host Configuration Protocol (DHCP)
Dynamic Host Configuration Protocol (DHCP)
General Information (DHCP)
The Dynamic Host Configuration Protocol (DHCP) is an auto configuration
protocol used on IP networks. The DHCP functionality is available on all
Agilent HPLC modules with on-board LAN Interface or LAN Interface Card,
and “B”-firmware (B.06.40 or above).
When the initialization mode “DHCP” is selected, the card tries to download
the parameters from a DHCP Server. The parameters obtained become the
active parameters immediately. They are not stored to the non-volatile
memory of the card.
Besides requesting the network parameters, the card also submits its
hostname to the DHCP Server. The hostname equals the MAC address of the
card, e.g.
0030d3177321
. It is the DHCP server's responsibility to forward the
hostname/address information to the Domain Name Server. The card does not
offer any services for hostname resolution (e.g. NetBIOS).
Figure 91
DHCP (Principle)
'+&3
6HUYHU
$FWLYH
3DUDPHWHU
N O T E
1
It may take some time until the DHCP server has updated the DNS server with the
hostname information.
2
It may be necessary to fully qualify the hostname with the DNS suffix, e.g.
0030d3177321.country.company.com
.
3
The DHCP server may reject the hostname proposed by the card and assign a name
following local naming conventions.
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......