![Agilent Technologies 1260 Infinity G1315C Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-g1315c/1260-infinity-g1315c_user-manual_2867535059.webp)
Agilent 1260 Infinity DAD and MWD User Manual
59
Installing the Module
3
Installing Capillaries
Installation of the Bio-inert Zero Dead Volume (ZDV) Union
The Bio-inert ZDV (p/n 5067-4741) union has two different connectors where
capillaries need to be installed in the correct sequence. Otherwise, an inset of
the union may be damaged and the connection may not be tight.
C A U T I O N
Potential leak or damage of the Bio-inert ZDV Union.
➔
To avoid leaks or a damage to the Bio-inert ZDV union, follow the procedure below
in the prescribed sequence.
1
Install the capillary at the end marked with a
ring/indentation.
2
Install the second capillary at the other end.
5LQJ
5LQJ
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......