52
Agilent 1260 Infinity DAD and MWD User Manual
3
Installing the Module
Flow Connections to the Detector
N O T E
The heat exchanger/capillary and the cell body can be fixed mirror symmetrically to have
both capillaries routed to the bottom or to the top (depending on the routing of the
capillaries to the column). For details see
“Replacing Capillaries on a Standard Flow
1
Press the release buttons and remove the front cover to
gain access to the flow cell area.
2
Press the release button and open the flow cell door.
3
Insert the flow cell.
4
Connect the flow cell capillaries to the capillary holder
(top is inlet, bottom is outlet).