![Agilent Technologies 1260 Infinity G1315C Скачать руководство пользователя страница 46](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-g1315c/1260-infinity-g1315c_user-manual_2867535046.webp)
46
Agilent 1260 Infinity DAD and MWD User Manual
3
Installing the Module
Installation Information on Leak and Waste Handling
1
Stack the modules according to the adequate stack configuration.
The leak pan outlet of the upper module must be vertically positioned above
the leak tray of the lower module, see
2
Connect data and power cables to the modules, see section
Installing the
Module
below.
3
Connect capillaries and tubes to the modules, see section
Flow Connections
to the module
below or the relevant system manual.
WA R N I N G
Toxic, flammable and hazardous solvents, samples and reagents
➔
Keep solvent path free from blockages.
➔
Keep the flow path closed (in case the pump in the system is equipped with a
passive inlet valve, solvent may leak out due to hydrostatic pressure, even if your
instrument is off).
➔
Avoid loops.
➔
Tubes must not sag.
➔
Do not bend tubes.
➔
Do not immerse tube end in waste liquid.
➔
Do not intubate tubes in other tubes.
➔
For correct tubing follow instructions on label attached to the module.
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......