Agilent 1260 Infinity DAD and MWD User Manual
83
Using the Detector
4
Special Settings of the Detector
Stop time
elapsed
•
Run in progress - Data Analysis
•
Prerun / Rawdata
•
Elapsed run time stops
•
Spectra counter continues
•
ChemStation continues to add missing data
Run ends
•
Ready
•
Run finished
•
Prerun / Ready
Table 12
Automated Run Recovery in case of temporary communication failures
Situation
Reaction
On ChemStation
N O T E
If the detector status window is not opened, you will realize only the Power Fail error and
the long Run In Progress information until the data is recovered from disk.
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......