Agilent 1260 Infinity DAD and MWD User Manual
5
How to optimize the Detector
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Slit Width
The detector has a variable slit at the entrance of the spectrograph. This is an
effective tool to adapt the detector to changing demand of different analytical
problems.
A narrow slit provides spectral resolution for analytes with very fine
structures in the absorbance spectrum. An example of such a spectrum is
benzene. The five main absorbance bands (fingers) are only 2.5 nm wide and
just 6 nm apart from each other.
Figure 38
Benzene at 1, 4 and 16 nm slit width (principle)
A wide slit uses more of the light shining through the flow cell. This gives
lower baseline noise as shown in
QP
QP
QP
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......