Agilent 1260 Infinity DAD and MWD User Manual
2
Site Requirements and Specifications
Performance Specifications
Flow cells
Standard: 13 µL volume, 10 mm cell
path length and 120 bar (1740 psi)
pressure maximum
Standard bio-inert: 13 µL volume,
10 mm cell path length and 120 bar
(1740 psi) pressure maximum
Semi-micro: 5 µL volume, 6 mm cell
path length and 120 bar (1740 psi)
pressure maximum
Micro: 2 µL volume, 3 mm cell path
length, 120 bar (1740 psi) pressure
maximum
Semi-nano: 500 nL volume, 10 mm cell
path length and 50 bar (725 psi)
pressure maximum
Nano: 80 nL volume, 6 mm cell path
length and 50 bar (725 psi) pressure
maximum
High pressure: 1.7 µL volume, 6 mm cell
path length and 400 bar (5800 psi)
pressure maximum
Prep SST: 3 mm cell path length and
120 bar (1740 psi) pressure maximum
Prep Quartz:0.3 mm cell path length
and 20 bar (290 psi) pressure maximum
Prep Quartz: 0.06 mm cell path length
and 20 bar (290 psi) pressure maximum
All flow cells are equipped with
RFID tags that hold cell typical
information.
pH range 1.0—9.5 (12.5 solvent
dependent with bio-inert version)
Time programmable
Wavelength, polarity, peak width, lamp
bandwidth, autobalance, wavelength
range, threshold, spectra storage mode
Spectral tools
Data analysis software for spectra
evaluation, including spectral libraries
and peak purity functions
Control and data
evaluation
Agilent ChemStation for LC (32-bit)
For 1260 systems:
•
Revision B.04.02 DSP2 or
above
For 1100/1200 systems:
•
Revision B.01.03 SR-2 /
B.02.01 SR-2 or above
Table 4
Performance Specifications G1315D
Type
Specification
Comments
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......