![Agilent Technologies 1260 Infinity G1315C Скачать руководство пользователя страница 101](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-g1315c/1260-infinity-g1315c_user-manual_2867535101.webp)
Agilent 1260 Infinity DAD and MWD User Manual
101
How to optimize the Detector
5
Optimization Overview
Optimization Overview
Table 15
Optimization Overview
Parameter
Impact
1
Selection of flow cell
•
peak resolution versus sensitivity
•
Choose flow cell according to used column, see
2
Connection of flow cell
•
chromatographic resolution
•
For flow rates from 0.5 mL/min connect column using the
zero-dead-volume fittings of the detector.
•
For small column i.d. (e.g 1 mm) the inlet capillary of the micro flow cell
can be connected directly to the column.
3
Setting the peak width (response time)
•
peak resolution versus sensitivity
versus disk space
•
Use peak width according
on page 102 as starting point.
•
Set the peak-width close to the width of a narrow peak of interest in your
chromatogram.
4
Setting wavelength and bandwidth
•
Sample wavelength:
•
Never miss a peak by the use of a browser wavelength like 250 nm
with 100 nm bandwidth.
•
Select specific wavelength with reduced bandwidth if you need
selectivity, e.g. 250,10 nm and 360,100 nm as reference wavelength.
•
Set the sample wavelength to a peak or valley in the spectrum to get
best linearity for high concentrations.
•
sensitivity versus selectivity
•
sensitivity versus linearity
•
Reference wavelength:
•
Select the reference wavelength with broad bandwidth (30...100 nm)
wavelength range where your analytes have little or no absorbance
(e.g. sample at 254 nm, reference at 320 nm).
•
baseline drift due to RI effects.
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......