![Agilent Technologies 1260 Infinity G1315C Скачать руководство пользователя страница 250](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-g1315c/1260-infinity-g1315c_user-manual_2867535250.webp)
250
Agilent 1260 Infinity DAD and MWD User Manual
12
Hardware Information
Interfaces
Interfaces Overview
CAN
The CAN is inter-module communication interface. It is a 2-wire serial bus
system supporting high speed data communication and real-time requirement.
LAN
The modules have either an interface slot for an LAN card (e.g. Agilent
G1369B/C LAN Interface) or they have an on-board LAN interface (e.g.
detectors G1315C/D DAD and G1365C/D MWD). This interface allows the
control of the module/system via a PC with the appropriate control software.
Some modules have neither on-board LAN nor an interface slot for a LAN card
(e.g. G1170A Valve Drive or G4227A Flex Cube). These are hosted modules and
require a Host module with firmware B.06.40 or later or with additional
G1369C LAN Card.
RS-232C (Serial)
The RS-232C connector is used to control the module from a computer
through RS-232C connection, using the appropriate software. This connector
can be configured with the configuration switch module at the rear of the
module. Refer to
Communication Settings for RS-232C
.
The RS-232C is designed as DCE (data communication equipment) with a
9-pin male SUB-D type connector. The pins are defined as:
N O T E
If an Agilent detector (DAD/MWD/FLD/VWD/RID) is in the system, the LAN should be
connected to the DAD/MWD/FLD/VWD/RID (due to higher data load). If no Agilent
detector is part of the system, the LAN interface should be installed in the pump or
autosampler.
N O T E
There is no configuration possible on main boards with on-board LAN. These are
pre-configured for
•
19200 baud,
•
8 data bit with no parity and
•
one start bit and one stop bit are always used (not selectable).
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......