Agilent 1260 Infinity DAD and MWD User Manual
173
Maintenance
9
Exchanging a Flow Cell
1
Press the release buttons and remove the front cover to
gain access to the flow cell area.
2
Open the flow cell cover.
N O T E
Depending on the system setup, the inlet capillary
might be routed directly from the module above or
below to the cell and not to the capillary holder.
3
Disconnect the flow cell inlet capillary (top) and the
waste tubing (bottom) from the unions.
4
Loosen the thumb screw and remove the flow cell outlet
capillary (bottom) with the union.
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......