Setting up an Analysis
1
Open the purge valve of your pump (by turning it counterclockwise) and set
flow rate to 3 – 5 mL/min.
2
Flush all tubes with at least 30 mL of solvent.
3
Set flow to required value of your application and close the purge valve.
Pump for approximately 10 minutes before starting your application.
Table 8
Choice of Priming Solvents for Different Purposes
Activity
Solvent
Comments
After an installation
When switching between reverse
phase and normal phase (both times)
Isopropanol
Isopropanol
Best solvent to flush air out of the
system
Best solvent to flush air out of the
system
After an installation
Ethanol or Methanol
Alternative to Isopropanol (second
choice) if no Isopropanol is available
Cleaning the system when using
buffers
After a solvent change
Bidistilled water
Bidistilled water
Best solvent to re-dissolve buffer
crystals
Best solvent to re-dissolve buffer
crystals
After the installation of normal phase
seals (P/N 0905-1420)
5% Isopropanol
Good wetting properties
N O T E
The pump should never be used for priming empty tubings (never let the pump run dry).
Use a syringe to draw enough solvent to completely fill the tubings up to the pump inlet
before you continue priming with the pump.
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......