Site Requirements and Specifications
2
Performance Specifications
Local Control
Agilent Instant Pilot (G4208A)
For 1260 systems:
•
B.02.11 or above
For other systems:
•
B.02.09 or above
Analog outputs
Recorder/integrator: 100 mV or 1 V,
output range 0.001 – 2 AU, two outputs
Communications
Controller-area network (CAN),
RS-232C, APG Remote: ready, start,
stop and shut-down signals, LAN
Safety and
maintenance
Extensive diagnostics, error detection
and display (through control module
and ChemStation), leak detection, safe
leak handling, leak output signal for
shutdown of pumping system. Low
voltages in major maintenance areas.
GLP features
RFID for electronics records of flow cell
and UV lamp conditions (path length,
volume, product number, serial number,
test passed, usage)
Early maintenance feedback (EMF) for
continuous tracking of instrument
usage in terms of lamp burn time with
user-setable limits and feedback
messages. Electronic records of
maintenance and errors. Verification of
wavelength accuracy with built-in
holmium oxide filter.
Housing
All materials recyclable.
Others
Electronic temperature control (ETC) for
the complete optical unit
Table 4
Performance Specifications G1315D
Type
Specification
Comments
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......