R3132 Series Spectrum Analyzer Operation Manual
5.2.19 Gain Compression
5-78
Instruments Required:
Procedures:
Initializing the power meter and the sensor
1.
On the power meter, perform the zeroing and calibration with power sensor.
2.
On the power meter, set the unit into dBm mode, after calibration has done.
Under preamplifier off condition
Setting the measurement conditions (Frequency 200 MHz)
Instruments Qty.
Recommended
Model
Signal Generator
1
SMP02
Signal Generator
1
SMP04
Power Meter
1
NRVS
Power Sensor
1
NRV-Z55
Power Splitter
1
DDUL-20A-1000
3 dB Attenuator
1
DEE-000685-1
20 dB Attenuator
1
DEE-000480-1
RF Cable SMA(m)-SMA(m)
3
A01002
Minimum Loss Pad
1
*1
HP11852B
Adapter N(m)-SMA(f)
2
Adapter SMA(f)-SMA(f)
1
*2
*1 R3132N Only
*2 R3182 Only