5.2.21 Residual Response
5-89
Under preamplifier on condition
13. On the R3132 series, press LEVEL and Hi Sens ON/OFF(ON) to set preampli-
fier to on.
14. On the R3132 series, press DISPLAY, Disp Line ON/OFF(ON), 1, 0, 5 and
MHz(-dBm) to place display line at -105 dBm position.
15. Repeat step 5 through 12.
Following steps are applied for R3162/72/82
Under preamplifier on condition
Measurement the residual response on the 3.2 GHz to 8 GHz for R3162
Measurement the residual response on the 3.2 GHz to 7.1 GHz for R3172/82
16. On the R3162/72/82, set controls as follows:
Center Frequency:
3.325 GHz
Span:
50 MHz
CF Step Size:
47.5 MHz
RBW:
100 kHz
VBW:
300 Hz
17. On the R3162/72/82, press DISPLAY, Disp Line ON/OFF(ON), 9, 0 and MHz
(-dBm) to place display line at -90 dBm position.
18. Repeat step 5 through 12 until the center frequency of 7.975 GHz (R3162) or
7.075 GHz (R3172/82).