R3132 Series Spectrum Analyzer Operation Manual
5.2.14 Accuracy of Frequency Readout and Count Frequency Marker
5-53
Measuring the Count Frequency Marker Accuracy
9.
On the signal generator, set controls as follow:
Output Level:
0 dBm
10. On the signal generator, set controls as follow:
Frequency:
2 GHz
11. On the R3132 series, after preset, set controls as follows:
Center Frequency:
2 GHz
Frequency Span:
1 MHz
12. On the R3132 series, press PK SRCH to capture the signal peak.
13. On the R3132 series, press as follows to set frequency counter mode to on.
COUNTER, Res 1Hz
14. On the R3132 series, press SINGLE for single sweep.
15. Record the frequency counter reading on the performance verification record
sheet.
16. Repeat steps 10 through 15 for each frequency setting listed in Table 5-19.
In step 11., the center frequency is more than 3.3 GHz on the R3162/72/82, press
as follows to tune preselector peak after the center frequency and the frequency
span setting.
REPEAT, PK SRCH, FREQ, 1/2_more, Presel, Auto Tune
Wait till auto tuning has completed.