R3132 Series Spectrum Analyzer Operation Manual
3.3.28 TRACE Key (Trace Data)
3-79
Average CONT/SGL
Toggles between CONT (continuation) and SGL (single) modes.
CONT:
Continues to average using the current data which is
used until the set averaging count is reached.
SGL:
Automatically switches to View mode as soon as the
desired averaging count has been reached.
When combining this function with OBW, ACP or
Spectrum Mask measurement, perform the averaging
process the number of times specified by this function
before making the OBW or ACP measurement.
Single Measure ON/OFF
Sets the sweep conditions used in Single Sweep Mode.
ON:
Pressing the SINGLE key performs the sweep the spec-
ified number of measurement times when the channel
power, total power, or average power is measured while
the average or power average function is turned on.
Note that the average and power average functions re-
quire the OBW, ACP (FULL, CARRIER), and spec-
trum mask measurement functions.
OFF:
Pressing the SINGLE key performs the sweep only
once.
Min Hold A(B)
Sets the Min Hold mode which is used to display the minimum
value for each trace sample (when you select the Min Hold mode,
the negative mode is automatically selected. Refer to Nega.)
Store A(B) to B(A)
Stores trace A(B) data as trace B(A).
Math A
Displays the Math A menu.
A-B
→
→
→
→
A
A: When trace A is in the Write mode, the contents of memory B
are subtracted from the sweep result and, then, the result is saved
in memory A.
When the Display line is active, the Display line value is added to
the value obtained previously and saved to A (A - B + DL
→
A).
B-A
→
→
→
→
A
A: When trace A is in the Write mode, the sweep result is sub-
tracted from the contents of memory B and, then, the result is
saved in memory A.
When the Display line is active, the Display line value is added to
the value obtained previously and saved to A (B - A + DL
→
A).
A-DL
→
→
→
→
A
A: When trace A is in the Write mode, the level of the display line
is subtracted from the sweep result and, then, the result is saved in
memory A.
Disp Line ON/OFF
Turns the display line, which is used as the reference line for com-
paring the trace level, on or off.