R3132 Series Spectrum Analyzer Operation Manual
5.2.16 Frequency Response
5-61
Instruments Required:
Procedures:
Setting the measurement conditions
1.
On the power meter, perform the zeroing and calibration with power sensor.
2.
On the power meter, set the unit into dBm mode, after calibration has done.
3.
On the power meter, set the correction frequency to 30 MHz.
4.
Connect the power meter, the signal generator and the power splitter as shown
Figure 5-16.
Figure 5-16 Setup of Frequency Response Test
Instruments Qty.
Recommended
Model
Signal Generator
1
SMP04
Power Meter
1
NRVS
Power Sensor
1
NRV-Z55
Power Splitter
1
K241C
Minimum Loss Pad
1
*1
HP11852B
RF Cable BNC(m)-BNC(m)
1
MI-09
RF Cable SMA(m)-SMA(m)
3
A01002
RF Cable K(m)-K(m)
1
*2
Adapter N(m)-SMA(f)
2
Adapter K(f)-K(f)
1
*2
*1 R3132N Only
*2 R3182 Only
Minimum
Loss Pad (*1)
*1 R3132N only