5.2.9 Input Attenuator Switching Accuracy (R3172/82)
5-32
5.2.9
Input Attenuator Switching Accuracy (R3172/82)
This test measures input attenuator’s switching accuracy over the full 70 dB.
The number of frequency measured points is three points at 4 GHz, 15 GHz and 18 GHz.
IF Gain uncertainty is measured when the resolution bandwidth is set to 1 kHz and result is filled in on the
IF Gain uncertainty of the performance verification record sheet.
Specification:
Reference to 10 dB input attenuation.
Apply for R3172
±
1.1 dB/10 dB step
Maximum deviation
±
2 dB
9 kHz to 12 GHz
±
1.3 dB/10 dB step
Maximum deviation
±
2.5 dB 12 GHz to 18 GHz
±
1.8 dB/10 dB step
Maximum deviation
±
3.5 dB 18 GHz to 26.5 GHz
Apply for R3182
±1.1 dB/10 dB step
Maximum deviation ±2 dB
9 kHz to 12 GHz
±1.3 dB/10 dB step
Maximum deviation ±2.5 dB 12 GHz to 18 GHz
±1.8 dB/10 dB step
Maximum deviation ±3.5 dB 18 GHz to 26.5 GHz
±2.2 dB/10 dB step
Maximum deviation ±4 dB
26.5 GHz to 40 GHz
Instruments Required:
Procedures:
Setting the measurement conditions
1.
Connect the step attenuators as shown in Figure 5-9.
Instruments
Qty.
Recommended Model
Signal Generator
1
SMIQ02E
1 dB Step Attenuator
1
HP8494H
10 dB Step Attenuator
1
HP8495H
Attenuator Driver
1
HP11713
RF Cable SMA(m)-SMA(m)
3
A01002
RF Cable BNC(m)-BNC(m)
1
MI-09
Adapter N(m)-SMA(f)
4
Adapter SMA(f)-SMA(f)
1
*
* R3182 Only