R3132 Series Spectrum Analyzer Operation Manual
5.2.20 Sweep Time Accuracy
5-85
Table 5-26 Sweep Time Accuracy Test Setting
For R3132 series
For signal generator
Sweep Time
Pulse Period
Pulse Width
50 sec
45 sec
1 sec
5 sec
4.5 sec
1 sec
500 msec
450 msec
200 msec
50 msec
45 msec
20 msec
5 msec (OPT29)
4.5 msec
2 msec
500
µ
sec (OPT29)
450
µ
sec
200
µ
sec
50
µ
sec (OPT29)
45
µ
sec
20
µ
sec