R3132 Series Spectrum Analyzer Operation Manual
2.3.1 Measuring the Channel Power
2-56
Setting the measurement conditions
This changes the analyzer settings so that the input signal is displayed more clearly.
5.
Press FREQ, 1, 9, 1, 7, ., 9, 5, 0 and MHz.
A center frequency of 1917.950 MHz is set.
6.
Press SPAN, 1 and MHz.
A frequency span of 1 MHz is set.
7.
Press BW, RBW AUTO/MNL(MNL), 1, 0 and kHz.
The RBW is set to 10 kHz.
8.
Press LEVEL, 1, 0 and GHz(+dBm).
The reference level is set to 10 dBm.
Setting the offset level
9.
Press 1/2_more, Ref Offset ON/OFF(ON), 1, 0 and GHz(dB).
The offset level is set to 10 dB.
The measurement values, including values for the external attenuator, are now
displayed.
Figure 2-49 Setting the Offset Level
Measuring the power
10. Press POWER MEASURE.
The Power menu is displayed.
11. Press Channel Power.
The CH Power menu is displayed.
12. Press Window Position, 1, 9, 1, 7, ., 9, 5, 0 and MHz.
The channel is set to 1917.950 MHz.