R3132 Series Spectrum Analyzer Operation Manual
7.2 R3132N Specifications
7-6
(2)
Amplitude Range
(3)
Sweep
Characteristics
Specification
Measurement range
+134 dB
µ
V to displayed Average Noise Level
Maximum input level
Preamplifier OFF
Preamplifier ON
(Input attenuator
≥
10 dB)
+134 dB
µ
V
±
50 VDC max.
+120 dB
µ
V
±
50 VDC max.
Display range
Log
Linear
10
×
10 div
10, 5, 2, 1 dB/div
10% of reference level/div
Reference level range
Preamplifier OFF
Log
Linear
Preamplifier ON
Log
Linear
(Input attenuator 0 to 50 dB)
+44.8 dB
µ
V to +148.8 dB
µ
V (0.1 dB step)
172.8
µ
V to +27.39 V
(Input attenuator 0 to 30 dB)
+26.8 dB
µ
V to +118.8 dB
µ
V (0.1 dB step)
21.75
µ
V to 866 mV
Input attenuator range
0 to 50 dB (5 dB step)
Characteristics
Specification
Sweep time
20ms to 1000s
Sweep time Accuracy
±
2%
Trigger mode
FREE RUN, LINE, VIDEO, EXT, TV
Sweep Mode