Chapter 3
Configuring settings
47
OsmoTECH XT Single-Sample Micro-Osmometer User Guide
7.
Press
SAVE
.
Each time a test is run, its results are automatically
prepared for export to the selected network folder.
3.10 Assigning user restrictions
Administrators can set which features supervisors and
operators can access before or after adding users to the
instrument.
If authorized by an administrator, supervisors can set
which features operators can access.
The features that supervisors and operators can be
given access are listed below.
9
Feature is accessible to these users depending on
restrictions setting.
8
Feature is never accessible to these users.
Feature
Supervisor
access
available
Operator
access
available
9
9
9
9
9
9
Calibration
9
9
9
9
Logging routine instrument cleaning
(sample probe and chamber)
9
9
9
9
9
9
Remove remote lock (see
)
9
9
9
8
(data
management)
9
8
Assigning user restrictions
9
8
(system
software/firmware)
9
8
1.
Press
SETTINGS > RESTRICTIONS
and log in.
2.
On the
USER ACCESS
tab, Administrators can enable/
disable supervisor and operators access to features.
If the
Change Restrictions
checkbox is selected,
supervisors can enable/disable the features that
operators can access.
3.
Press
SAVE
.
Changing password settings
Administrators can change the password strength
requirements for user passwords, such as:
•
Required length (number of characters)
•
Minimum number of lower case/UPPER case
characters
•
Minimum number numeric characters.
•
Minimum number of special characters (#, ~, !, etc.)
1.
Press
SETTINGS >
RESTRICTIONS > PASSWORDS
.
2.
Set passwords to expire at 30 days, 60 days,
90 days, 180 days, or never.
3.
Select the
Pre-expire passwords on creation
checkbox to make users change their password the
first time they log in.