Appendix E
OsmoTECH XT LIS Specification
87
OsmoTECH XT Single-Sample Micro-Osmometer User Guide
8.4.12
Action code (optional)—Single character
indicating the action to be taken when the order
is received. See the definition of the codes in the
LIS2-A2 document.
8.4.15
Date and time specimen received (optional)—Text
field indicating the date/time of receiving the
order.
8.4.16
Specimen descriptor (optional)—Text field
indicating the type of the specimen.
8.4.17
Ordering physician (optional)—Text field
indicating the name of the ordering physician.
Parts of the name should be separated with
component delimiters rather than space
characters.
8.4.18
Physician’s telephone number (optional)—Text
field indicating the telephone number of the
ordering physician.
8.4.19
User field number 1 (optional)—Text field sent by
the requester and returned with the response.
8.4.21
Laboratory field number 1—Text field definable for
any use by the laboratory. This field is optional.
8.4.23
Date and time results reported—Text field
indicating the date/time that the order was
reported back with the result. The instrument
always includes this field in reports sent to the
LIS.
8.4.26
Report type (optional)— Single character
specifying the type of the report. See the
definition of the codes in the LIS2-A2 document.
Example
O|1|3MA005||^^^||20161110082002|
20161110082002||0.000|||||
20161110082002||||||||20161110082002|
||<CR><ETX>
E.9 Result record
The fields used in this record are as specified in LIS2-A2:
9.1
Record type ID—
R
identifies the record as a result
record.
9.2
Sequence number—Sequence number (Integer)
of the result record associated with the preceding
test order record.
9.3
Universal test ID—Text field containing up to four
parts describing the test. This code is always the
four-letter code
OSMO
for the instrument.
9.4
Measured value—Number (Integer) indicating the
osmolality value.
9.5
Units—Text field indicating the units for the
preceding value. This is always
mOsm/Kg H2O
for the instrument.
9.9
Result status (optional)—Text field containing the
status of the test. See the LIS2-A2 document for
the possible codes.
9.11
Operator identification (optional, UTF8-
encoded)—Text field that identifies the operator
who performed the test and possibly also the test
verifier, separated by a component delimiter.
9.12
Date/time of start of test—Text field containing
the date/time the test was started.
9.13
Date/time of end of test—Text field containing
the date/time the test finished.
9.14
Instrument Identification—Text field containing
the instrument serial number.
Example
R|1|^^^OSMO|51|mOsm/Kg H2O||N|N|F|
|OperatorID|20161027142723|
|17010095A<CR><ETX>