![Advanced Instruments OsmoTECH XT Скачать руководство пользователя страница 8](http://html.mh-extra.com/html/advanced-instruments/osmotech-xt/osmotech-xt_user-manual_2856049008.webp)
Quality control
9
OsmoTECH XT Single-Sample Micro-Osmometer User Guide
Interpreting results
Data generated using Advanced Instruments standards and reference solutions may be analyzed according to the accuracy
and precision specifications of the instrument (see
Laboratories may choose to employ one, two, or three standard deviations (SD) for accuracy based on what is relevant in
their laboratories. For normally distributed data, approximately 68% of the individual data values will fall within one
standard deviation of the mean, approximately 95% within two standard deviations, and approximately 99.7% within three
standard deviations.
Advanced Instruments Standards
and Reference Solutions
Precision
Accuracy (mOsm/kg H
2
O)
1 SD
2 SD
3 SD
Clinitrol™ 290 Reference Solution
Standard deviation
≤
2 mOsm/kg H
2
O
288–292
286–294
284–296
0 mOsm/kg Calibration Standard
Standard deviation
≤
2 mOsm/kg H
2
O
0-2
0-4
0-6
50 mOsm/kg Calibration Standard
Standard deviation
≤
2 mOsm/kg H
2
O
48–52
46–54
44–56
100 mOsm/kg Calibration Standard
Standard deviation
≤
2 mOsm/kg H
2
O
98–102
96–104
94–106
200 mOsm/kg Calibration Standard
Standard deviation
≤
2 mOsm/kg H
2
O
198–202
196–204
194–206
300 mOsm/kg Calibration Standard
Standard deviation
≤
2 mOsm/kg H2O
298–302
296–304
294–306
400 mOsm/kg Calibration Standard
Standard deviation
≤
2 mOsm/kg H
2
O
398–402
396–404
394–406
500 mOsm/kg Calibration Standard
Coefficient of variation
≤
0.5%
497-503
494-506
491-509
850 mOsm/kg Calibration Standard
Coefficient of variation
≤
0.5%
845-855
840-860
835-865
900 mOsm/kg Calibration Standard
Coefficient of variation
≤
0.5%
895-905
890-910
885-915
1000 mOsm/kg Calibration Standard
Coefficient of variation
≤
0.5%
995–1005
990–1010
985–1015
1500 mOsm/kg Calibration Standard
Coefficient of variation
≤
1%
1485–1515
1470–1530
1455–1545
2000 mOsm/kg Calibration Standard
Coefficient of variation
≤
1%
1980–2020
1960–2040
1940–2060
3000 mOsm/kg Calibration Standard
Coefficient of variation
≤
1%
2970–3030
2940–3060
2910–3090
4000 mOsm/kg Calibration Standard
Coefficient of variation
≤
1%
3960–4040
3920–4080
3880–4120