Osmometer Model 3900 User’s Guide
“System Error:
Requests”
Primitive”
Interrupt”
Communica-
tions”
Unknown”
Trap”
NMI”
ID”
“Test Time-out
Error”
“Thermoelectri
cs Driver
Failure”
“Truncated
Barcode”
“WARNING-
sample Count
Incorrect”
Switch the instrument off,
then on.
Check the sample probe
by running the A/D Test.
1. Was the sample OK?
2. Is the block probe bin #
correct?
3. Check the alignment by
running the Head
Up/Down Test.
Switch the instrument off,
then on.
Truncated bar codes
longer than 16 characters
may appear as duplicate
entries.
Produced by a recall oper-
ation or at the completion
of a batch/bar code run if
# samples recorded does
not match that entered for
batch/bar code run.
Symptom
Check
Suggested Action
If the error message does
not persist and other mes-
sages are not displayed,
ignore this message.
Replace the sample probe.
1. Prepare another sample
and repeat the test.
2. Check/correct the block
probe bin number.
3. If the problem persists,
see Chapter 5 for further
assistance.
Retry the test. If the prob-
lem persists, see Chapter 5
for further assistance.
Try to avoid the use of bar
codes longer than 16 char-
acters.
1. Check original printout
for # entered vs. # run.
2. If test errors occurred,
these may not have been
recorded as valid tests,
resulting in the message.
3. If test errors add up to
the # entered, disregard.
Содержание 3900
Страница 1: ...The Advanced Osmometer Model 3900 User s Guide 3905 Rev20 061213 ...
Страница 12: ...xii The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 64: ...46 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 68: ...50 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 114: ...96 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 118: ...The Advanced Osmometer Model 3900 User s Guide 100 Notes ...
Страница 124: ...106 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 128: ...The Advanced Osmometer Model 3900 User s Guide 110 Date Problem Symptom Action ...