92
Performance Over
Operating Conditions
Temperature
Less than 1 mOsm/kg H
2
O for every 5°C (9°F) ambient
Effects:
temperature change.
Operating
Conditions
Temperature:
Low Range: 18°C to 35°C (64°F to 95°F)
High Range: 18°C to 30°C (64°F to 86°F)
Humidity:
5 to 80% relative humidity (non-condensing)
Storage Temp.:
-40°C to +45°C (-40°F to +113°F)
Start-up Time:
Actual time from power-on to “Osmometer Ready” varies
according to ambient temperature and humidity -- typically 3
to 10 minutes. Maximum throughput may require up to 30
minutes of equilibration time from power-on.
Test Time:
90 seconds per sample (typical)
Sample
Approximately 30 samples per hour at 23°C (73°F); less at
Throughput:
higher ambient temperature and in high range
Dimensions
inches
centimeters
Width:
21.5
54.6
Depth:
20.0
50.8
Height:
21.5
54.6
Weight
pounds
kilograms
Net:
66.0
30.0
Shipping:
105.0
47.7
Warranty:
One-year limited warranty on workmanship and all parts
except glass, plastic, and parts warranted by their makers
Certification:
Refer to Regulatory Notices (Appendix C) for applicable
standards
Installation Class:
I
Over-Voltage
Category:
II
Pollution Degree:
2
Moisture Protection:
IPX0 (ordinary)
Содержание 3900
Страница 1: ...The Advanced Osmometer Model 3900 User s Guide 3905 Rev20 061213 ...
Страница 12: ...xii The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 64: ...46 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 68: ...50 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 114: ...96 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 118: ...The Advanced Osmometer Model 3900 User s Guide 100 Notes ...
Страница 124: ...106 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 128: ...The Advanced Osmometer Model 3900 User s Guide 110 Date Problem Symptom Action ...