cassette to the back left corner of the left (input) deck and press the
START button.
When finished testing, press STOP to exit from the test menu to
“
Osmometer Ready
”.
Adjustment and replacement of parts and
assemblies
1. Air Filter Replacement
The air filters (figure 2) should be cleaned or replaced monthly or
whenever they become clogged with dust to the point where samples
cool more slowly than normal. To change or clean the air filters:
a. Remove the filters by gripping the edge of the square finger-hole
and pulling straight out.
b. With the open sides of the finger-holes down, push the new air
filters firmly, straight in as far as they will go.
2. Fuse Replacement
The power switch and fuse holder are located beside the power cord
connector on the back panel of the instrument. Switch the power
switch to the off ( O ) position and disconnect the power cord. Use
a small flat-bladed screwdriver or similar tool to pry open the fuse
holder door. Visually check for a blown fuse. If there is any doubt,
test the fuses with a continuity checker or ohmmeter, or simply
replace them with new fuses. Double check the values marked on
the fuses. The 3900 will automatically adjust for voltages between
100VAC and 250VAC, but appropriately rated fuses must be
installed. Use 5 × 20mm, 250V, time delay (Type T) fuses. For 100-
130V operation, use 5-Amp fuses; for 200-250V operation, use 2.5-
Amp fuses.
67
Troubleshooting & Service
Содержание 3900
Страница 1: ...The Advanced Osmometer Model 3900 User s Guide 3905 Rev20 061213 ...
Страница 12: ...xii The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 64: ...46 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 68: ...50 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 114: ...96 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 118: ...The Advanced Osmometer Model 3900 User s Guide 100 Notes ...
Страница 124: ...106 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 128: ...The Advanced Osmometer Model 3900 User s Guide 110 Date Problem Symptom Action ...