89
“Sample Probe
Failure”
“Sample Probe
Open?”
“Sample Tube
Error”
“Standards
Reversed?
Repeat Calib”
“Stir/Freeze
Driver Failure”
“Supervisor
Key Needed;
Press STOP” or
“Supervisor
Key Needed”
Switch the instrument off,
then on.
Check the sample probe
by running the A/D Test.
Switch the instrument off,
then on.
Check the sample probe
by running the A/D Test.
Check for a sample tube
in the wrong place, possi-
bly in the operating head
or freezing chamber.
Review procedure and
retry the test.
Did you introduce the
right calibration standards
in the right order?
Switch the instrument off,
then on.
Check the key switch
position.
Symptom
Check
Suggested Action
If the error message does
not persist and other mes-
sages are not displayed,
ignore this message.
Replace the sample probe.
If the error message does
not persist and other mes-
sages are not displayed,
ignore this message.
Replace the sample probe.
Remove the misplaced
tube, check the probe and
stir/freeze wire alignment
and retry the test.
If the problem persists, see
Chapter 5 for further assis-
tance.
Recalibrate with the cor-
rect calibration standards.
Retry the test. If the prob-
lem persists, see Chapter 5
for further assistance.
Place the key switch in the
supervisor position before
attempting to use CALIB,
TEST or SETUP.
Troubleshooting Table
Содержание 3900
Страница 1: ...The Advanced Osmometer Model 3900 User s Guide 3905 Rev20 061213 ...
Страница 12: ...xii The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 64: ...46 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 68: ...50 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 114: ...96 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 118: ...The Advanced Osmometer Model 3900 User s Guide 100 Notes ...
Страница 124: ...106 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 128: ...The Advanced Osmometer Model 3900 User s Guide 110 Date Problem Symptom Action ...