The Advanced
®
Osmometer Model 3900 User’s Guide
84
“Parameter
RAM Failed”
or “New
Software
Version”
“RAM Failure”
“Recalibration
Needed”
“Reset Probe
Configuration”
displayed dur-
ing the power-
up sequence
“Reset Probe
Configuration”
displayed dur-
ing the power-
up sequence,
followed by
“Select Test
Item”
Switch the instrument off,
then on.
Calibrate the instrument.
Symptom
Check
Suggested Action
If not followed by “Reset
Probe Configuration”, see
Chapter 5 for further assis-
tance.
If followed by “Reset
Probe Configuration”, see
below.
If the problem persists, see
Chapter 5 for further assis-
tance.
If the error message
repeats after successful re-
calibration, see Chapter 5
for further assistance.
If not followed by “Select
Test Item”, see Chapter 5
for further assistance.
If followed by “Select Test
Item”, see below.
1. Press > to advance to
“Set Block Probe Bin #”
in the SETUP menu and
reset the block probe bin
#.
2. Press > to advance to
“Set Sample Bin #” in
the SETUP menu and
reset the sample probe
bin #.
(continued...)
Содержание 3900
Страница 1: ...The Advanced Osmometer Model 3900 User s Guide 3905 Rev20 061213 ...
Страница 12: ...xii The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 64: ...46 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 68: ...50 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 114: ...96 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 118: ...The Advanced Osmometer Model 3900 User s Guide 100 Notes ...
Страница 124: ...106 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 128: ...The Advanced Osmometer Model 3900 User s Guide 110 Date Problem Symptom Action ...