79
Troubleshooting Table
Appendix A
Abrupt loss
of power (no
display)
1. Power availability.
Temporarily unplug the
power cord from its wall
outlet and replace it with
a known workable lamp
or small appliance.
2. Power connection.
3. Unplug the power cord
from the power cord
socket. Use a small
screwdriver to pry out
the voltage selector/fuse
holder. Visually check
the two 5 mm x 20 mm
time delay (T) fuses
mounted on it. Use a
continuity checker, or
replacement fuse to
check continuity if there
is any doubt.
Check the values
marked on the fuses. For
100-130V operation
they should be 2-Amp
fuses; for 200-250V
operation they should be
1-Amp fuses.
Symptom
Check
Suggested Action
1. Have the power restored.
2. Make sure the power
cord is firmly plugged
into both the wall outlet
and the socket on the
back of the instrument.
3. Replace the fuses if
blown or if incorrect
value.
Reinstall the fuse holder
into the back of the
instrument with the cor-
rect arrowheads match-
ing.
If the fuses were blown
and the replacement fuses
blow also, see Chapter 5
for further assistance.
Содержание 3900
Страница 1: ...The Advanced Osmometer Model 3900 User s Guide 3905 Rev20 061213 ...
Страница 12: ...xii The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 64: ...46 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 68: ...50 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 114: ...96 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 118: ...The Advanced Osmometer Model 3900 User s Guide 100 Notes ...
Страница 124: ...106 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 128: ...The Advanced Osmometer Model 3900 User s Guide 110 Date Problem Symptom Action ...