8. Press START. The Model 3900 will position the cassette, wipe the
probe and stir wire between the probe wipers, pick up the first sam-
ple tube, place it in the sample freezing chamber and perform the
freezing point test.
The sample temperature will be displayed as soon as it falls below
0°C. The Model 3900 will freeze the sample at the crystallization
point speci fied by the range and “buzz” point selection, follow the
plateau development, determine the osmolality and display and
record the results. After completing the test, the Model 3900 will
return the first sample tube to the cassette, wipe the probe and return
to the cassette for the second tube. Figure 8 shows the operating
head either loading or unloading the second sample tube from a cas-
sette.
35
Figure 8:
Operating Head Loading
or Unloading a Sample Tube
Instrument Operation