Sample Preparation
No further sample preparation is required. Body fluids such as serum
or plasma may be used directly.
NOTE
Additives such as anticoagulants from collection tubes may
contribute signifi cantly to the measured osmolality.
Particulate matter can cause premature crystallization; in blood it is nor-
mally eliminated by centrifugation. Particulate matter in urine may be
removed by centrifu gation or filtration through an inert filter.
Sample test procedure
If the Model 3900 has been turned off, it should be started as indicated
in Installation & Setup.
The default range is 0-1550 mOsm and the default “buzz” point is
3000 mOsm. If the expected freezing points are in the 1450-4000
mOsm range and the op tional high range has been implemented, the
high range should be selected before beginning a set of tests. If the
freezing point of a sample is not within the range selected, a message
will be dis played, requesting the range be changed.
Like most measuring instruments, osmometers need calibration against
standards. Your Model 3900 was calibrated at the factory using the pro-
cedures described in Calibration. The probe and calibration parameters
are stored in parameter RAM which is powered by an internal bat tery
when the instrument power is off or disconnected. Thus, when
“
Osmometer Ready
” is displayed, your Model 3900 is calibrated and
ready to run. However, the calibration should always be verified during
opera tion to be sure that none of the parameters affecting calibration
have changed. In the recommended proce dure which follows, calibra-
tion verification is accom plished by starting with the Clinitrol™ refer-
ence solu tion. Then, if recalibration is required, see Calibration.
It is important that if you are running in low range you select the cor-
rect sample tube type you are using, (glass/plastic). Use the range
32
The Advanced
®
Osmometer Model 3900 User’s Guide
Содержание 3900
Страница 1: ...The Advanced Osmometer Model 3900 User s Guide 3905 Rev20 061213 ...
Страница 12: ...xii The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 64: ...46 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 68: ...50 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 114: ...96 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 118: ...The Advanced Osmometer Model 3900 User s Guide 100 Notes ...
Страница 124: ...106 The Advanced Osmometer Model 3900 User s Guide Notes ...
Страница 128: ...The Advanced Osmometer Model 3900 User s Guide 110 Date Problem Symptom Action ...