7451A/7461A/7461P Digital Multimeter Operation Manual
7.1.3 Resistance Measurement (2W
, LP-2W
, 4W
, LP-4W
)
7-8
Additional error depending on the integration time
Voltage between input terminals without connections.9.5 V or less
Maximum allowable applied voltage
•
7451A
2W
and 4W
measurements
Integration time
Additional error
(digit + m
)
10
s
IT
200
s
-
In a 3 1/2-digit display
2+20
200
s < IT
2 ms
Includes FAST1 and FAST2. In a 4 1/2-digit display
2+20
2 ms <
IT < 1 PLC
-
In a 5 1/2-digit display
2+20
1 PLC
IT < 10 PLC
Integral multiple of 1 PLC
In a 6 1/2-digit display
1+0
10 PLC < IT
10 s
Integral multiple of 1 PLC
In a 6 1/2-digit display
2+0
Between
and COM
1000 Vpeak
Between COM and the chassis
500 V
Between 4W
Hi-4W
Lo
200 Vpeak
Between 4W
Hi, 4W
Lo and the chassis
500 V
Range
Maximum
display
Resolution
Measurement
current
Measurement accuracy *1
(% of r digits)
Temperature coefficient *5
(ppm of r digits)/
C
24 hours
23
C
1
C *2
90 days
23
C
5
C
1 year
23
C
5
C
Auto-zero
ON
Auto-zero
OFF
30
31.9999
100
10 mA
0.0030+30.0
0.0100+40.0
0.0150+40.0
10.0+5.0
10.0+50.0
300
319.999
1 m
1 mA
0.0020+5.0
0.0080+11.0
0.0150+11.0
10.0+1.0
10.0+6.0
3000
3199.99
10 m
1 mA
0.0020+3.0
0.0070+3.0
0.0120+3.0
10.0+0.2
10.0+3.0
30 k
31.9999
100 m
100
A
0.0020+3.0
0.0070+3.0
0.0130+3.0
10.0+0.2
10.0+3.0
300 k
319.999
1
10
A
0.0020+3.0
0.0090+3.0
0.0140+3.0
8.0+0.2
8.0+3.0
3000 k
3199.99
10
1
A
0.0070+14.0
0.0300+19.0
0.0300+19.0
35.0+5.0
35.0+5.0
30 M
31.9999
100
100 nA
0.0600+14.0
0.1800+19.0
0.2000+19.0
100.0+5.0 *3
100.0+5.0 *3
300 M
319.999
1 k
10 nA
0.6000+14.0
1.7000+19.0
2.0000+19.0
1000.0+5.0 *4 1000.0+5.0 *4
*1
Integration time: 10 PLC, Auto-zero: ON for 2W
, Display digit: 5 1/2
The measurement cable resistance and 0.2
(internal circuit resistance error) are added to the digit error in 2W
measurement.
*2
Relative value with respect to the calibration standard
*3
115 ppm/
C is added to the temperature coefficient in the temperature range of 30
C to 50
C.
*4
1150 ppm/
C is added to the temperature coefficient in the temperature range of 30
C to 50
C.
*5
The temperature coefficient in Auto-zero ON is used in the 4W
measurement.