7.1.8 Continuity Measurement
7-16
7.1.8
Continuity Measurement
•
7461A/61P
•
7451A
7.1.9
Temperature Measurement (TEMP-2W, TEMP-3W, TEMP-4W)
7461P
Platinum resistance bulb
Pt100 (compliant with JIS C1604-1997) or
JPt100 (compliant with JIS C1604-1981)
Measurement range
PT100
-200
C to +850
C
JPt100
-200
C to +649
C
Linearization
Digital calculation
Wire connection
2W, 3W or 4W
Allowable wire resistance 10
or lower (except 2W)
Measurement unit
C,
F or K
Range
Maximum display
Resolution
Measurement current
1000
1199.999
1 m
1 mA
Specifications not described above are the same as those in the 1000
range in the 2W
measurement.
Continuity judgment setting range: 1
to 1000
Range
Maximum display
Resolution
Measurement current
3000
3199.99
10 m
1 mA
Specifications not described above are the same as those in the 3000
range in the 2W
measurement.
Continuity judgment setting range: 1
to 1000
Resolution
Measurement
current
Measurement accuracy
(% of r
C)*1
Temperature coefficient
(ppm of r
C)/
C *1
90 days, 23
5
C
1 year, 23
5
C
0.001
C
1 mA
0.006 + 0.06
0.008 + 0.06
6 + 0.003
*1 4W, RATE: SLOW2, not including the measurement probe accuracy