7451A/7461A/7461P Digital Multimeter Operation Manual
4.2.1 Measurement Functions
4-9
Figure 4-6 AC Voltage Measurement (ACV and ACV(AC+DC))
4. Resistance measurement (2W
, 4W
, LP-2W
, and LP-4W
)
1. Press .
The resistance is measured by using the 2W
or 4W
which is set before this
measurement.
2. Press
and
.
The resistance is measured in low-power in which the measurement current is
reduced.
3. Press
and
to return to the 2W
and 4W
measurement.
4. Press .
The 2W
and 4W
are switched alternately.
NOTE:
In the following measurement ranges, the LO terminal of 4W
SENSE does
not recognize the voltage and operates as in 2W
measurement.
7461A: 1000
k
range or more in 4W
, 100 k
range or more in
LP-4W
7451A: 3000
k
range or more in 4W
, 300 k
range or more in
LP-4W
Table 4-7 Measurement range
7461A/61P
7451A
Measurement method
TrueRMS
Measurement range
5% or more of the range
Crest factor
5
3
1000Vpeak
AC700Vrms or less
2.2
10
7
VHz
500V or less
SHIFT
SHIFT