7451A/7461A/7461P Digital Multimeter Operation Manual
6.5 Comparator Output
6-12
The calculation result signals of Hi, Go, and Lo are output to the signal output pins of HI, PASS, and LO.
The output to the PASS pin depends on the PASS conditions as shown in Table 6-3 and Table 6-4.
Table 6-3 Optical Semiconductor Relay Contact and TTL Logical Output (a)
Judgment result
Contact/TTL output pin
RELAY HI
RELAY PASS/TTL
PASS
RELAY LO
TTL FAIL
Hi
make
break
Go
break
break
Lo
break
make
Table 6-4 Optical Semiconductor Relay Contact and TTL Logical Output (b)
PASS conditions
RELAY PASS
TTL PASS
TTL FAIL
Judgment result
Judgment result
Panel setting
Remote Command
(ADC command)
Lo
Go
Hi
Lo
Go
Hi
LGH
LOP1
MIP1
HIP1
L
L
L
H
H
H
--H
LOP0
MIP0
HIP1
H
H
L
L
L
H
-G-
LOP0
MIP1
HIP0
H
L
H
L
H
L
L--
LOP1
MIP0
HIP0
L
H
H
H
L
L
L-H
LOP1
MIP0
HIP1
L
H
L
H
L
H
LG-
LOP1
MIP1
HIP0
L
L
H
H
H
L
-GH
LOP0
MIP1
HIP1
H
L
L
L
H
H
---
LOP0
MIP0
HIP0
H
H
H
L
L
L
H: break or Hi, L: make or Low