![ADCMT 7451A Скачать руководство пользователя страница 180](http://html1.mh-extra.com/html/adcmt/7451a/7451a_operation-manual_2848273180.webp)
7451A/7461A/7461P Digital Multimeter Operation Manual
7.1.10 Measurement Time and Display Digits
7-18
Integration Time (IT) and Sampling Interval (SI)
Integration Time and Display Digits
7461A/61P
7451A
Integration time (IT)
Integration time (IT)
Setting range
Resolution
Setting range
Resolution
10
s to 199
s
1
s 100
s to 199
s
1
s
200
s to 9.99 ms
10
s 200
s to 9.99 ms
10
s
10 ms to 5999.9 ms
100
s 10 ms to 5999.9 ms
100
s
6 s to 10 s
200
s 6 s to 10 s
200
s
0.02 PLC to 100 PLC
0.01 PLC 0.02 PLC to 100 PLC
0.01 PLC
Sampling interval (SI)
Sampling interval (SI)
Setting range
Setting range
50
s to 3600 s
200
s to 3600 s
Function
DCV/DCI
DCV-Bch
2W/4W
LP-2W/4W
ACV/ACI
Diode and Continuity
ACV(AC+DC)
ACI(AC+DC)
Integration time (IT)
7461A/61P
7451A
7461A/61P
7451A
10
s
(IT)
200
s
3 1/2-digit display 3 1/2-digit display
3 1/2-digit display
200
s < (IT)
2 ms
4 1/2-digit display 4 1/2-digit display
3 1/2-digit display
2 ms < (IT)
15 ms
5 1/2-digit display 5 1/2-digit display
4 1/2-digit display
15 ms < (IT)
10 s
6 1/2-digit display 5 1/2-digit display
4 1/2-digit display