7451A/7461A/7461P Digital Multimeter Operation Manual
6.7.3 ADC Command Reference
6-23
Measurement
DCV-Bch
measurement mode
* Valid only in the
DCV-Bch func-
tion (F12)
SDV0
HI-COM measurement
SDV1
HI-LO measurement
SDV2
RATIO measurement
SDV?
Reply: One of SDV0 to SDV2
Invalid function
INH?n
Reads the invalid state of the function.
n: 1 (DC voltage measurement) to 50 (Frequency
measurement)
Reply 0: Valid state of the function selection
1: Invalid state of the function selection
Trigger
*TRG
Trigger command
Measurement data
memory
ST0
Store: OFF
ST1
Store: ON
ST?
Reply: ST0 or ST1
IRDn, m Readout range setting
n, m: 0 to 9999 (7451A/61A)
0 to 19999 (7461P)
• Parameters cannot be omitted.
(0, 0)
IRO?
Data reading
Reply: See Section 6.7.2, “Output Data Format.”
IRPO?
Number of measurement results readings
Reply: IRPOdddd (7451A/61A)
IRPOddddd (7461P)
IRNO?
Data range reading
Reply: IRNOdddd, dddd (7451A/61A)
IRNO
ddddd,
ddddd (7461P)
No data: IRNO0000, -001 (7451A/61A)
IRNO+00000, -00001 (7461P)
ICL
Measurement data memory initialization
Trigger system
Start
INI
Leaves the IDLE state.
Continue
INIC0
CONTINUOUS OFF
INIC1
CONTINUOUS ON
INIC?
Reply: INIC0 or INIC1
Trigger abort
ABO
Moves to the IDLE state.
Trigger source
selection
TRS0
IMMEDIATE
TRS1
MANUAL
TRS2
EXTERNAL
TRS3
BUS
TRS4
LEVEL
TRS5
DELTA
TRS?
Reply: One of TRS0 to TRS5
Item
Command
Description
Initial value