5 Operations After Start of Probing
5-11
[Lot management setting]
[Map input setting]
Perform Lot Management?
Choose "Yes".
# of Cassette Per Lot
Enter the number of cassettes of the lot.
Processing When Lot # Not Input
Select "Alarm Stop" except when lot number can be
identified with the wafer ID.
Clear Lot # At Lot End?
Select "Yes" except when lot number can be
identified with the wafer ID.
Touching "
Combination of Wafer Management ID
"
will open the special selection screen as shown at left.
Choose an appropriate combination.
Input Map Data?
Choose "Yes".
Input Map Data From:
Choose "FD".
Input Format of Map Data
Select "TSK Standard".