![ACCRETECH UF190R Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/accretech/uf190r/uf190r_user-manual_3046036076.webp)
4 Wafer Mapping & Die Sampling
4-16
For the sample die testing, open [Mapping Settings]
sub-menu Page 2/2 and touch “
Sampling Test Die
Settings
”; the display changes to [Sampling Test Die
Settings] menu as shown below.
DATA IN At Sampling Start Die?
This parameter is used to select the specification of
sampling start die in the lot first wafer. If “Yes” is
selected, data-in request of the start die appears on
the first wafer. If “No” is selected, probing of each
registered sample die is immediately started.
Sampling Die Setting
“Die by Die” is for Data-in on each sampling die; for a
multi-site probing Data-in is done on Ch. 1 die.
“Block” is to choose blocks of die for sampling test
(see the figure below), applicable at the single site
probing only.
“Down Loading” is not effective at present.
Sampling Die Selection
With the top “DATA IN At Sampling Start Die” set to
‘No’, choose ‘Absolute Coordinates’, when the sample
die locations are fixed on the wafer map for each
wafer.
With “DATA IN At Sampling Start Die” set to ‘Yes’,
choose ‘Relative Coordinates’.
With “Sampling Die Setting” set to “Relative
Coordinates”, the coordinate value changes by the
Data-in position of the sampling start die on the lot
first wafer.
Sampling Start Die Coordinates X/Y:
Input range: -511 ~ 511
Starting die for sampling measurement is specified.
Block Size for Block Sampling X/Y
The die arraying directions from the Data-in die of the
block are determined by the probing directions. (1 ~
20)
Data-in die
Size Y
Size X