5 Operations After Start of Probing
5-3
5.2 Overdrive Amount Modification
To change the probe tip overdrive amount during lot
processing, first stop the probing and then touch the
<Needle Align. Posi. Correct> button on the text map
or Prober status screen.
The display changes to the contact height manual
adjustment screen (shown on left). Touch the
<Overdrive Change> button; the screen again
changes to the [Probing Parameter] menu which
contains the parameter “Probing Overdrive” alone, as
shown below.
Other buttons on the contact height manual
adjustment screen:
<Image Display Change>
Touch this button twice to move the chuck to below
the alignment camera and switch the display to the
probe mark checking screen. (Refer to the following
page.)
<CANCEL>
Stops the contact height determination process
started with the <Needle Align. Posi. Correct>
button.
<Data-in>
If the contact height determination function is not
available, you can readjust the probe tip contact
height while observing the probe tips via the
camera. Touch the <Data-in> button after adjusting
the Chuck height with the <Z-UP/DOWN> switch.