3 Automatic Needle Alignment
3-19
DEVICE : TEST
INDEX : 4410.0 8830.0
ORI-FLA : 180
XY LIMIT(1): 20 (from pad edge)
XY LIMIT(2): 30 (from pad center)
Z LIMIT : 20
No. X Y DX DY DZ ERR
01-001 +01000 +01000 +5 +10 0
01-002 +01500 +01000 +40 +5 10 X
01-003 +02000 +01000 +15 +20 0
01-004 +02500 +01000 +10 –35 30 YZ
CARD CENTER : X= 00000.0 Y= 00000.0
CONTACT HEIGHT : Z= 17350
WAFER THICKNESS: 500
CARD ANGLE : 1.00000 deg
(Size/distance in µm)
X
Y
X
Y
Pad
X
Y
Card
CARD ANGLE
CARD
CENTER
Pushing <Print> button on the confirmation screen
makes the various data of the alignment result
printed out by the built-in or external printer as
shown at left. (With the Basic Operational
parameter "Output of Needle Alignment Results"
under [Setup Sequence Settings] - [Needle
Alignment Settings] set to "1: Yes" or "2: Yes (Only
Auto Focus Needle)", these data will be printed out
automatically at the correct finish of the pad
alignment processes. See Page 3-4.)
DEVICE :Device
name
INDEX
:Index sizes X, Y
ORI-FLA
:Orientation flat angle
XY LIMIT (1) :Tip position allowance from the pad
edge
XY LIMIT (2) :Tip position allowance from the pad
center
Z LIMIT
:Tip height allowance
No.
:Site No. - Pad No.
X, Y
:Pad center from the die lower left
street cross
DX, DY
:Tip position from the pad center
DZ
:Tip height from the average
ERR :Out-of-tolerance
items
CARD CENTER
:Card center displacement
CONTACT HEIGHT :Probe-pad contact height from
the Chuck origin (including the
wafer thickness)
WAFER THICKNESS: Thickness of wafers
CARD ANGLE :Card attachment angle
The figures shown left represent relationships
between the pad center and the die lower left street
cross and between the card center/card angle and
Prober X/Y axes. The card attachment angle is
taken from the X axis in the right direction, plus in
CCW, viewed from the top. For DX, DY, and DZ,
see Page 3-14.
The pad and probe tip count printed in "No." section
depends on "All pad", "Meas. pad" or "NG pad" in
the "Mode" selection.