NEUROPROBE
®
500 PRO USER MANUAL 29
COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
I. NERVE BLOCK PROBE TREATMENT PROTOCOL
The Nerve Block modality contains the following pre-programmed probe protocol (Probe Nerve Block):
PROBE NERVE BLOCK:
PARAMETER
DEFAULT VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
5.0 KHz
Stim time:
60 Seconds
0
– 999 Seconds
Waveform:
Alternating Current
Output:
0 mA
0
– 2.25 mA
II. NERVE BLOCK PADS AND SLD INFRARED PAD PROTOCOL
The Nerve Block modality contains the following pre-programmed Pads protocol (Pads Nerve Block):
PADS NERVE BLOCK:
PARAMETER
DEFAULT VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
5.0 KHz
Treatment time:
15 Minutes
0
– 99 Minutes
Auto intensity:
0%
0
– 20%
Waveform:
Alternating Current
Output:
0 mA
0
– 70 mA