Press the MODIFY TIME

 

/

 

PARAMETER button to set the desired treatment time if you wish to change 

from  the default  settings of 15  minutes. Press  the up or down  button  to increase or decrease the duration 
from a minimum of one minute to a maximum of 99 minutes.   The timer will count down to zero during 
treatment. 

NOTE:

 

The SLD Infrared Pads always operate for the indicated treatment time for the selected protocol 

with  the  same  output  regardless  of  the  electrotherapy  protocol  chosen.  The  SLD  Infrared  Pads  pulse  at 
sequential frequencies of 45.4, 62.5, 83.3, KHz at a 50% duty factor. Each frequency is active for 1 minute. 
The  frequency  sequence  is  repeated  until  the  timer  terminates  the  treatment.  This  sequencing  is  not 
discernable to the user. 

 

To stop the treatment, push the START/STOP button (3) once. 

 

When the treatment time is down to zero, the treatment current is switched off and INFRARED THERAPY is 
displayed. 

 

Turn off the unit by pressing the POWER button (1/5).

 

 

IFC Therapy with Pad Electrodes 

 

 

Press the POWER button (1/5) to turn on the Neuroprobe

®

 500 Pro. The display screen will illuminate, and 

the unit will be in the SELECT

 

MODALITY/PROGRAM mode. 

 

Select the treatment modality desired and the appropriate treatment program by using the modality buttons 
(2) to select Sensory Motor, or Nerve Block pad programs. 

 

Press  the  SELECT  PARAMETERS  button  to  select  any  settings  to  be  modified.    Press  the  MODIFY 
TIME/PARAMETER button to adjust the desired treatment time, or other settings, if you wish to change from 
the default  settings.  Time  is preset  to 15 or 30  minutes (dependent on the protocol  selected). Press the  up or 
down buttons to increase or decrease the duration from a minimum of one minute to a maximum of 99 minutes. 

 

Position the electrodes on the patient, connect the electrode lead wires and press the START/STOP

 

button 

(3).  The treatment timer will begin, and the screen will provide basic treatment parameters, and will read 
as follows (i.e. if Pads Nerve Block was selected): 

 

 

 

 

 

 

 

 

 

 

 

 

 

IFC FF    5.0 KHz                                     T15 
 
A: 00 mA     B: 00 mA

                    

 

Содержание 140500B

Страница 1: ......

Страница 2: ...al may not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual without prior review and written authoriz...

Страница 3: ...rapeutic Ultrasound Pulsed Shortwave Diathermy devices and advanced Therapeutic Exercise Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNICYCLE OMNIVR OMNISTAND OMNIVERSA OMNISWD and SYNCHRONY...

Страница 4: ...cuments Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust acciden...

Страница 5: ...STIMULATION Optional 15 TREATMENT GUIDELINES 16 Introduction to Medium Frequency Currents 16 Electrode Application Techniques 17 SLD Infrared Pad Application Techniques 19 Treatment Preparation 19 INF...

Страница 6: ...frared spectrum to provide topical heating for the purpose of elevating tissue temperature for Temporary relief of minor muscle and joint pains Minor arthritis pain and muscle spasm Relieving stiffnes...

Страница 7: ...ximity less than 5 feet to an operating shortwave or microwave diathermy unit may produce instability in the device output or burns at the treatment site Lead wires and device can pick up the magnetic...

Страница 8: ...s are not advised for use with OmniVersa Omnistim Systems When cleaning the device never immerse them or wash them with water See the infection control section in this manual for cleaning instructions...

Страница 9: ...n Establishment of acceptable intensity levels for desensitized areas may be related to the intensity levels tolerated on normal skin in opposite or related body parts Caution should be used in the pr...

Страница 10: ...on of an Estim probe Interferential Current in bipolar or quadripolar modes or Monochromatic Infrared Therapy to reduce pain and increase circulation Flexible neoprene pads deliver heat and Infrared t...

Страница 11: ...selected Time remaining Carrier Frequency Bar graph for Vector and detection modes Button speeds and delay Vector selection Output current Display contrast Sweep rate and modulation selection E LEAD...

Страница 12: ...output or stop the treatment START STOP also cancels the SET mode Q LED for Point Stimulation Probe output indicator R LED indicating unit is connected to line power Factory Settings The Neuroprobe 5...

Страница 13: ...xtremity SLD Infrared Pad Plug it into the probe output socket Apply the SLD Infrared Pad to the treatment area Use an infection control cover over the SLD Infrared Pad and proper infection control pr...

Страница 14: ...he treatment current is switched off and INFRARED THERAPY is displayed Turn off the unit by pressing the POWER button 1 5 IFC Therapy with Pad Electrodes Press the POWER button 1 5 to turn on the Neur...

Страница 15: ...Note that for all IFC or INFRARED PAD protocols the treatment time is preset but may be adjusted at will during the treatment Once the treatment is completed remove the electrodes from the patient an...

Страница 16: ...probe or from the output A control on the panel The stim light on the probe turns on when the unit is in stimulation mode and shuts off during detection mode The treatment is complete when all of the...

Страница 17: ...r or in Quadripolar Interferential mode for patient treatment With medium frequency currents the energy of each individual pulse is low providing for stimulation of only one or two neurons Since the p...

Страница 18: ...arious areas of the body Electrode placement is dependent on the etiology of the condition Monopolar Mono Polar Technique This technique may use two electrodes of different sizes however one is in the...

Страница 19: ...other CERVICAL BIPOLAR PLACEMENT OF ELECTRODES IFC Quadripolar Quadri Polar Technique This technique requires the use of two output channels and four electrodes usually but not exclusively of the same...

Страница 20: ...n indication of increased current density under part of or the entire electrode surface In this case immediately start to reduce the current intensity to zero remove the electrodes to inspect the surf...

Страница 21: ...gelled or self adhering electrodes all increase the risk of irritation or burns and are contraindicated for use with Neuroprobe Systems Lead Wires Inspect the full length of the lead wires for signs...

Страница 22: ...sure to blood body fluids containing blood semen and vaginal secretions is likely to occur Health care workers with exudating lesions or weeping dermatitis should refrain from all direct patient care...

Страница 23: ...reased risk of patient cross contamination The following are the recommended practices After each use clean common contact surfaces such as control panel and lead wires with ACP germicidal wipes With...

Страница 24: ...assessment tools cart handles etc 7 Set up the patient per guidelines for the procedure 8 Provide treatment as appropriate 9 Discard all disposables 10 With clean gloves remove the plastic film from...

Страница 25: ...PADS NERVE BLOCK Symptomatic relief of pain from localized dermal or segmental origin 5000Hz Continuous Treatment time of 15 minutes Block pain by causing temporary nerve block via reactive depolariza...

Страница 26: ...e and the red wires in the opposite phase relative to each channel All variable parameters may be altered during operation I SENSORY MOTOR PROBE TREATMENT PROTOCOLS The Sensory Motor protocols contain...

Страница 27: ...onal electrotherapy approaches Its primary application is in the reduction of pain and in the stimulation of increased blood flow in the deeper tissues and muscles The following pads programs are avai...

Страница 28: ...0 seconds 0 20 Seconds Beat Sweep Rate Scan Time Modulation ON ON or OFF Vector Auto Intensity 0 0 20 Treatment time 30 minutes 0 99 Minutes Waveform Burst Alternating Current Output 0 mA 0 99 mA PADS...

Страница 29: ...y be varied during operation and saved by the user The following program is available in Noxious Probe mode PROBE NOXIOUS Sensory Noxious Waveform PARAMETER DEFAULT VALUES RANGE OF USER ADJUSTABLE VAL...

Страница 30: ...USER ADJUSTABLE VALUES Carrier frequency 5 0 KHz Stim time 60 Seconds 0 999 Seconds Waveform Alternating Current Output 0 mA 0 2 25 mA II NERVE BLOCK PADS AND SLD INFRARED PAD PROTOCOL The Nerve Block...

Страница 31: ...if the lead wire passes the test LED will be red if the lead wires are faulty and need to be replaced see Figure 7 below Remove electrode s and replace if necessary Adjust output up or down as necess...

Страница 32: ...d INSPECT POWER SUPPLY OPERATION 1 Connect the power cord to the power supply and to a power outlet 2 Connect the power supply cable to the stimulator unit Figure 4 Figure 4 3 Press the power supply s...

Страница 33: ...ceptacle labeled LEAD WIRE TESTER When the POWER is on green LED will turn on if the lead wire passes the test LED will be red if the lead wires are faulty and need to be replaced Figure 7 Service Cen...

Страница 34: ...S RESERVED For repair or service of ACP Products and accessories please call 800 350 1100 and follow the prompts Normal hours of operation are 6 00am to 5 00pm Pacific Standard Time After hours please...

Страница 35: ...r Scans amplitude of channel A from 0 maximum every 10 seconds relative to channel B which varies concurrently out of phase Slow 90 Scans amplitude of channel A from 0 maximum every 60 seconds relativ...

Страница 36: ...conductivity PATIENT SAFETY SYSTEMS Activation By patient safety hand control which shuts down output Output modality may not be changed during operation Output levels are reset to zero at the start a...

Страница 37: ...ptional probe IFC and SLD Infrared Pad treatments Shipping Weight 4 lbs 1 8 kg ITEM ITEM NO DESCRIPTION 48758 Lead Wire Blue Green Coded plug in 7 214cm 080 tip pin stereo plug 55055 Grounding Pad 3x4...

Страница 38: ...stim Electrodes 4ea pkg 10pkg bx 61227 2x4 Multi use E stim Electrodes 4ea pkg 10pkg bx 26854 3x5 Multi use E stim Electrode 2ea pkg 10pkg bx SINGLE USE ELECTRODES 28303 2x2 Single use E stim Electrod...

Страница 39: ...trol of Touch Operatory Surfaces 4 x 6 Clear Perforated Sheets 1200 sheets roll 50593 Barrier Film for Infection Control of Infrared Therapy Pads 12 x 14 Perforated Sheets 800 sheets roll 63574 Barrie...

Страница 40: ...oid if any of the following have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been t...

Страница 41: ...ed repair is deemed to be under warranty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipp...

Страница 42: ......

Отзывы: