NEUROPROBE

®

 500 PRO USER MANUAL 

 

11 

 

COPYRIGHT 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED 

CONFIDENTIAL AND PROPRIETARY 

J.

 

INFRARED  PAD  OR  ESTIM  PROBE

 

connection.  Output  connector  for  the  Point  Stimulation  Probe  or  SLD 

Infrared Therapy Pads. 

K.

 

MODALITY SELECTION

 

buttons. Buttons select modalities as follows: 

a.

 

Selects INFRARED THERAPY mode when pressed. 

b.

 

Selects SENSORY MOTOR protocols when pressed. 

c.

 

Selects PROBE NOXIOUS protocols when pressed. 

d.

 

Selects NERVE BLOCK protocols when pressed. 

L.

 

DETECT

 

SENS button. Press the up arrow (

) button to increase detection sensitivity for measurement of dry 

skin. Press the down arrow (

) to decrease the detection sensitivity for measurement of wet or damp skin. 

M.

 

OUTPUT

 

A/B

 

buttons. Press the up arrow (

) button to increase output current.  Press the down arrow (

) to 

decrease output current. Select channel by color code displayed.  Channels are color coded (output buttons, lead 
wire receptacles, and lead wires are all coded green or blue) for convenient identification. 

N.

 

START/STOP button. Having selected the type of treatment and program, press to start the output or stop the 
treatment.  START/STOP

 

also cancels the SET

 

mode. 

O.

 

OUTPUT button. Use to increase/decrease the stimulation output in Point Stimulation Probe mode. 

P.

 

S

TART

/S

TOP

  button.  Having  selected  the  type  of  point  stimulation  treatment  and  program,  press  to  start  the 

output or stop the treatment.  START/STOP

 

also cancels the SET

 

mode. 

Q.

 

LED for Point Stimulation Probe output indicator

R.

 

LED

 

indicating unit is connected to line power.

 

 

Factory Settings 

 

The Neuroprobe

®

 500 Pro comes with the following factory settings: 

 

Screen Contrast 

Button Speed, A & B Output 

90 

Button Speed, Beat etc. 

90 

Button Delay 

20 

Audio 

ON 

LASER 

OFF 

 
These are generally suitable for most clinicians. To adjust these default factory settings proceed as follows: 
 
1.

 

When the  main  screen states  “SELECT

 

MODALITY/PROGRAM,”  push  the  SELECT/PARAMETER  button 

until the setting to be changed is displayed. The factory settings can then be adjusted by pressing the MODIFY 
TIME/PARAMETER button. 

2.

 

The  Screen  Contrast  sets  the  appropriate  viewing  angle  for  easy  screen  legibility.    Range  is  0  to  30,  with  30 
being the darkest display. 

3.

 

The Button Speed for Output A & B sets the speed at which output will increase and decrease when the buttons 
are pressed.  The range is 0 to 99 with 99 being the fastest. 

4.

 

The  Button  Speed,  BEAT  etc.  controls  the  scrolling  speed  of  all  of  the  remaining  buttons  on  the  system.    A 
speed of 90 to 95 is typical if the operator has good coordination and hand control.  The range is 0 to 99 with 99 
being the fastest. 

5.

 

The Button Delay represents the amount of time a button has to be pushed and held down before they will begin 
to  automatically  increment  or  decrement  at  the  button  speed. The  smaller  the  number,  the  faster  the  response 
time.  Range is 0 to 99. 

6.

 

Audio ON or OFF disables or enables the audio system which beeps at the end of the treatment. 

Содержание 140500B

Страница 1: ......

Страница 2: ...al may not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual without prior review and written authoriz...

Страница 3: ...rapeutic Ultrasound Pulsed Shortwave Diathermy devices and advanced Therapeutic Exercise Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNICYCLE OMNIVR OMNISTAND OMNIVERSA OMNISWD and SYNCHRONY...

Страница 4: ...cuments Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust acciden...

Страница 5: ...STIMULATION Optional 15 TREATMENT GUIDELINES 16 Introduction to Medium Frequency Currents 16 Electrode Application Techniques 17 SLD Infrared Pad Application Techniques 19 Treatment Preparation 19 INF...

Страница 6: ...frared spectrum to provide topical heating for the purpose of elevating tissue temperature for Temporary relief of minor muscle and joint pains Minor arthritis pain and muscle spasm Relieving stiffnes...

Страница 7: ...ximity less than 5 feet to an operating shortwave or microwave diathermy unit may produce instability in the device output or burns at the treatment site Lead wires and device can pick up the magnetic...

Страница 8: ...s are not advised for use with OmniVersa Omnistim Systems When cleaning the device never immerse them or wash them with water See the infection control section in this manual for cleaning instructions...

Страница 9: ...n Establishment of acceptable intensity levels for desensitized areas may be related to the intensity levels tolerated on normal skin in opposite or related body parts Caution should be used in the pr...

Страница 10: ...on of an Estim probe Interferential Current in bipolar or quadripolar modes or Monochromatic Infrared Therapy to reduce pain and increase circulation Flexible neoprene pads deliver heat and Infrared t...

Страница 11: ...selected Time remaining Carrier Frequency Bar graph for Vector and detection modes Button speeds and delay Vector selection Output current Display contrast Sweep rate and modulation selection E LEAD...

Страница 12: ...output or stop the treatment START STOP also cancels the SET mode Q LED for Point Stimulation Probe output indicator R LED indicating unit is connected to line power Factory Settings The Neuroprobe 5...

Страница 13: ...xtremity SLD Infrared Pad Plug it into the probe output socket Apply the SLD Infrared Pad to the treatment area Use an infection control cover over the SLD Infrared Pad and proper infection control pr...

Страница 14: ...he treatment current is switched off and INFRARED THERAPY is displayed Turn off the unit by pressing the POWER button 1 5 IFC Therapy with Pad Electrodes Press the POWER button 1 5 to turn on the Neur...

Страница 15: ...Note that for all IFC or INFRARED PAD protocols the treatment time is preset but may be adjusted at will during the treatment Once the treatment is completed remove the electrodes from the patient an...

Страница 16: ...probe or from the output A control on the panel The stim light on the probe turns on when the unit is in stimulation mode and shuts off during detection mode The treatment is complete when all of the...

Страница 17: ...r or in Quadripolar Interferential mode for patient treatment With medium frequency currents the energy of each individual pulse is low providing for stimulation of only one or two neurons Since the p...

Страница 18: ...arious areas of the body Electrode placement is dependent on the etiology of the condition Monopolar Mono Polar Technique This technique may use two electrodes of different sizes however one is in the...

Страница 19: ...other CERVICAL BIPOLAR PLACEMENT OF ELECTRODES IFC Quadripolar Quadri Polar Technique This technique requires the use of two output channels and four electrodes usually but not exclusively of the same...

Страница 20: ...n indication of increased current density under part of or the entire electrode surface In this case immediately start to reduce the current intensity to zero remove the electrodes to inspect the surf...

Страница 21: ...gelled or self adhering electrodes all increase the risk of irritation or burns and are contraindicated for use with Neuroprobe Systems Lead Wires Inspect the full length of the lead wires for signs...

Страница 22: ...sure to blood body fluids containing blood semen and vaginal secretions is likely to occur Health care workers with exudating lesions or weeping dermatitis should refrain from all direct patient care...

Страница 23: ...reased risk of patient cross contamination The following are the recommended practices After each use clean common contact surfaces such as control panel and lead wires with ACP germicidal wipes With...

Страница 24: ...assessment tools cart handles etc 7 Set up the patient per guidelines for the procedure 8 Provide treatment as appropriate 9 Discard all disposables 10 With clean gloves remove the plastic film from...

Страница 25: ...PADS NERVE BLOCK Symptomatic relief of pain from localized dermal or segmental origin 5000Hz Continuous Treatment time of 15 minutes Block pain by causing temporary nerve block via reactive depolariza...

Страница 26: ...e and the red wires in the opposite phase relative to each channel All variable parameters may be altered during operation I SENSORY MOTOR PROBE TREATMENT PROTOCOLS The Sensory Motor protocols contain...

Страница 27: ...onal electrotherapy approaches Its primary application is in the reduction of pain and in the stimulation of increased blood flow in the deeper tissues and muscles The following pads programs are avai...

Страница 28: ...0 seconds 0 20 Seconds Beat Sweep Rate Scan Time Modulation ON ON or OFF Vector Auto Intensity 0 0 20 Treatment time 30 minutes 0 99 Minutes Waveform Burst Alternating Current Output 0 mA 0 99 mA PADS...

Страница 29: ...y be varied during operation and saved by the user The following program is available in Noxious Probe mode PROBE NOXIOUS Sensory Noxious Waveform PARAMETER DEFAULT VALUES RANGE OF USER ADJUSTABLE VAL...

Страница 30: ...USER ADJUSTABLE VALUES Carrier frequency 5 0 KHz Stim time 60 Seconds 0 999 Seconds Waveform Alternating Current Output 0 mA 0 2 25 mA II NERVE BLOCK PADS AND SLD INFRARED PAD PROTOCOL The Nerve Block...

Страница 31: ...if the lead wire passes the test LED will be red if the lead wires are faulty and need to be replaced see Figure 7 below Remove electrode s and replace if necessary Adjust output up or down as necess...

Страница 32: ...d INSPECT POWER SUPPLY OPERATION 1 Connect the power cord to the power supply and to a power outlet 2 Connect the power supply cable to the stimulator unit Figure 4 Figure 4 3 Press the power supply s...

Страница 33: ...ceptacle labeled LEAD WIRE TESTER When the POWER is on green LED will turn on if the lead wire passes the test LED will be red if the lead wires are faulty and need to be replaced Figure 7 Service Cen...

Страница 34: ...S RESERVED For repair or service of ACP Products and accessories please call 800 350 1100 and follow the prompts Normal hours of operation are 6 00am to 5 00pm Pacific Standard Time After hours please...

Страница 35: ...r Scans amplitude of channel A from 0 maximum every 10 seconds relative to channel B which varies concurrently out of phase Slow 90 Scans amplitude of channel A from 0 maximum every 60 seconds relativ...

Страница 36: ...conductivity PATIENT SAFETY SYSTEMS Activation By patient safety hand control which shuts down output Output modality may not be changed during operation Output levels are reset to zero at the start a...

Страница 37: ...ptional probe IFC and SLD Infrared Pad treatments Shipping Weight 4 lbs 1 8 kg ITEM ITEM NO DESCRIPTION 48758 Lead Wire Blue Green Coded plug in 7 214cm 080 tip pin stereo plug 55055 Grounding Pad 3x4...

Страница 38: ...stim Electrodes 4ea pkg 10pkg bx 61227 2x4 Multi use E stim Electrodes 4ea pkg 10pkg bx 26854 3x5 Multi use E stim Electrode 2ea pkg 10pkg bx SINGLE USE ELECTRODES 28303 2x2 Single use E stim Electrod...

Страница 39: ...trol of Touch Operatory Surfaces 4 x 6 Clear Perforated Sheets 1200 sheets roll 50593 Barrier Film for Infection Control of Infrared Therapy Pads 12 x 14 Perforated Sheets 800 sheets roll 63574 Barrie...

Страница 40: ...oid if any of the following have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been t...

Страница 41: ...ed repair is deemed to be under warranty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipp...

Страница 42: ......

Отзывы: