NEUROPROBE

®

 500 PRO USER MANUAL 

 

COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED 

ELECTROTHERAPY/INFRARED INDICATIONS & CONTRAINDICATIONS 

 
CAUTION:  Federal  law  restricts  this  device  for  sale  or  use  by  or  on  the  order  of  a  Practitioner 
licensed by the laws of the state in which he/she practices to use or order the use of the device. 
 
Please  notice  that  Accelerated  Care  Plus  cannot  provide  medical  advice.  If  you  have  specific 
medical questions, please contact your healthcare professional. 
 

Indications 

Electrotherapy: 
Transcutaneous Electrical Nerve Stimulation (TENS) 

– Interferential Current (IFC) 

 

Symptomatic  relief  and  management  of  chronic  intractable  pain  and  as  an  adjunctive  treatment  in  the 
management of acute pain, post-surgical pain and pain associated with post-traumatic injury. 

 

Infrared: 
Super Luminous Diode (SLD) Infrared Therapy 

The SLD Infrared Therapy Pads are indicated for emitting energy in the infrared spectrum to provide topical heating 
for the purpose of elevating tissue temperature for: 

 

Temporary relief of minor muscle and joint pains. 

 

Minor arthritis pain and muscle spasm. 

 

Relieving stiffness. 

 

Promoting the relaxation of muscle tissue. 

 

Temporarily increase of local blood circulation. 

 

Contraindications  

Electrotherapy: 
Transcutaneous Electrical Nerve Stimulation (TENS) 

– Interferential Current (IFC) 

 

Transcutaneous electrical stimulation, or powered  muscle  stimulators should not be used on patients  with 
cardiac demand pacemakers and/or implanted defibrillators. 

 

Never  connect  lead  wires  to  the  power  line  or  electro-surgery  equipment.    Use  only  the  lead  wires 
recommended or approved by the manufacturer. 

NOTE:

  There  is  no  contraindication  to  the  application  of  Transcutaneous  Electrical  Stimulation  over  metal 

implants. 

 

Infrared: 
Super Luminous Diode (SLD) Infrared Therapy 

 

None. 

NOTE:

 The SLD Infrared Pads may be used on patients with implanted Pacemakers and electronic devices, as 

the  Infrared  therapy  will  not  affect  their  operations.  There  is  no  contraindication  to  the  application  of  SLD 
Infrared Pads over metal implants. 

Содержание 140500B

Страница 1: ......

Страница 2: ...al may not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual without prior review and written authoriz...

Страница 3: ...rapeutic Ultrasound Pulsed Shortwave Diathermy devices and advanced Therapeutic Exercise Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNICYCLE OMNIVR OMNISTAND OMNIVERSA OMNISWD and SYNCHRONY...

Страница 4: ...cuments Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust acciden...

Страница 5: ...STIMULATION Optional 15 TREATMENT GUIDELINES 16 Introduction to Medium Frequency Currents 16 Electrode Application Techniques 17 SLD Infrared Pad Application Techniques 19 Treatment Preparation 19 INF...

Страница 6: ...frared spectrum to provide topical heating for the purpose of elevating tissue temperature for Temporary relief of minor muscle and joint pains Minor arthritis pain and muscle spasm Relieving stiffnes...

Страница 7: ...ximity less than 5 feet to an operating shortwave or microwave diathermy unit may produce instability in the device output or burns at the treatment site Lead wires and device can pick up the magnetic...

Страница 8: ...s are not advised for use with OmniVersa Omnistim Systems When cleaning the device never immerse them or wash them with water See the infection control section in this manual for cleaning instructions...

Страница 9: ...n Establishment of acceptable intensity levels for desensitized areas may be related to the intensity levels tolerated on normal skin in opposite or related body parts Caution should be used in the pr...

Страница 10: ...on of an Estim probe Interferential Current in bipolar or quadripolar modes or Monochromatic Infrared Therapy to reduce pain and increase circulation Flexible neoprene pads deliver heat and Infrared t...

Страница 11: ...selected Time remaining Carrier Frequency Bar graph for Vector and detection modes Button speeds and delay Vector selection Output current Display contrast Sweep rate and modulation selection E LEAD...

Страница 12: ...output or stop the treatment START STOP also cancels the SET mode Q LED for Point Stimulation Probe output indicator R LED indicating unit is connected to line power Factory Settings The Neuroprobe 5...

Страница 13: ...xtremity SLD Infrared Pad Plug it into the probe output socket Apply the SLD Infrared Pad to the treatment area Use an infection control cover over the SLD Infrared Pad and proper infection control pr...

Страница 14: ...he treatment current is switched off and INFRARED THERAPY is displayed Turn off the unit by pressing the POWER button 1 5 IFC Therapy with Pad Electrodes Press the POWER button 1 5 to turn on the Neur...

Страница 15: ...Note that for all IFC or INFRARED PAD protocols the treatment time is preset but may be adjusted at will during the treatment Once the treatment is completed remove the electrodes from the patient an...

Страница 16: ...probe or from the output A control on the panel The stim light on the probe turns on when the unit is in stimulation mode and shuts off during detection mode The treatment is complete when all of the...

Страница 17: ...r or in Quadripolar Interferential mode for patient treatment With medium frequency currents the energy of each individual pulse is low providing for stimulation of only one or two neurons Since the p...

Страница 18: ...arious areas of the body Electrode placement is dependent on the etiology of the condition Monopolar Mono Polar Technique This technique may use two electrodes of different sizes however one is in the...

Страница 19: ...other CERVICAL BIPOLAR PLACEMENT OF ELECTRODES IFC Quadripolar Quadri Polar Technique This technique requires the use of two output channels and four electrodes usually but not exclusively of the same...

Страница 20: ...n indication of increased current density under part of or the entire electrode surface In this case immediately start to reduce the current intensity to zero remove the electrodes to inspect the surf...

Страница 21: ...gelled or self adhering electrodes all increase the risk of irritation or burns and are contraindicated for use with Neuroprobe Systems Lead Wires Inspect the full length of the lead wires for signs...

Страница 22: ...sure to blood body fluids containing blood semen and vaginal secretions is likely to occur Health care workers with exudating lesions or weeping dermatitis should refrain from all direct patient care...

Страница 23: ...reased risk of patient cross contamination The following are the recommended practices After each use clean common contact surfaces such as control panel and lead wires with ACP germicidal wipes With...

Страница 24: ...assessment tools cart handles etc 7 Set up the patient per guidelines for the procedure 8 Provide treatment as appropriate 9 Discard all disposables 10 With clean gloves remove the plastic film from...

Страница 25: ...PADS NERVE BLOCK Symptomatic relief of pain from localized dermal or segmental origin 5000Hz Continuous Treatment time of 15 minutes Block pain by causing temporary nerve block via reactive depolariza...

Страница 26: ...e and the red wires in the opposite phase relative to each channel All variable parameters may be altered during operation I SENSORY MOTOR PROBE TREATMENT PROTOCOLS The Sensory Motor protocols contain...

Страница 27: ...onal electrotherapy approaches Its primary application is in the reduction of pain and in the stimulation of increased blood flow in the deeper tissues and muscles The following pads programs are avai...

Страница 28: ...0 seconds 0 20 Seconds Beat Sweep Rate Scan Time Modulation ON ON or OFF Vector Auto Intensity 0 0 20 Treatment time 30 minutes 0 99 Minutes Waveform Burst Alternating Current Output 0 mA 0 99 mA PADS...

Страница 29: ...y be varied during operation and saved by the user The following program is available in Noxious Probe mode PROBE NOXIOUS Sensory Noxious Waveform PARAMETER DEFAULT VALUES RANGE OF USER ADJUSTABLE VAL...

Страница 30: ...USER ADJUSTABLE VALUES Carrier frequency 5 0 KHz Stim time 60 Seconds 0 999 Seconds Waveform Alternating Current Output 0 mA 0 2 25 mA II NERVE BLOCK PADS AND SLD INFRARED PAD PROTOCOL The Nerve Block...

Страница 31: ...if the lead wire passes the test LED will be red if the lead wires are faulty and need to be replaced see Figure 7 below Remove electrode s and replace if necessary Adjust output up or down as necess...

Страница 32: ...d INSPECT POWER SUPPLY OPERATION 1 Connect the power cord to the power supply and to a power outlet 2 Connect the power supply cable to the stimulator unit Figure 4 Figure 4 3 Press the power supply s...

Страница 33: ...ceptacle labeled LEAD WIRE TESTER When the POWER is on green LED will turn on if the lead wire passes the test LED will be red if the lead wires are faulty and need to be replaced Figure 7 Service Cen...

Страница 34: ...S RESERVED For repair or service of ACP Products and accessories please call 800 350 1100 and follow the prompts Normal hours of operation are 6 00am to 5 00pm Pacific Standard Time After hours please...

Страница 35: ...r Scans amplitude of channel A from 0 maximum every 10 seconds relative to channel B which varies concurrently out of phase Slow 90 Scans amplitude of channel A from 0 maximum every 60 seconds relativ...

Страница 36: ...conductivity PATIENT SAFETY SYSTEMS Activation By patient safety hand control which shuts down output Output modality may not be changed during operation Output levels are reset to zero at the start a...

Страница 37: ...ptional probe IFC and SLD Infrared Pad treatments Shipping Weight 4 lbs 1 8 kg ITEM ITEM NO DESCRIPTION 48758 Lead Wire Blue Green Coded plug in 7 214cm 080 tip pin stereo plug 55055 Grounding Pad 3x4...

Страница 38: ...stim Electrodes 4ea pkg 10pkg bx 61227 2x4 Multi use E stim Electrodes 4ea pkg 10pkg bx 26854 3x5 Multi use E stim Electrode 2ea pkg 10pkg bx SINGLE USE ELECTRODES 28303 2x2 Single use E stim Electrod...

Страница 39: ...trol of Touch Operatory Surfaces 4 x 6 Clear Perforated Sheets 1200 sheets roll 50593 Barrier Film for Infection Control of Infrared Therapy Pads 12 x 14 Perforated Sheets 800 sheets roll 63574 Barrie...

Страница 40: ...oid if any of the following have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been t...

Страница 41: ...ed repair is deemed to be under warranty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipp...

Страница 42: ......

Отзывы: