32
NEUROPROBE
®
500 PRO USER MANUAL
COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Figure 6
LEAD WIRE TESTER OPERATION
Lead Wires should be inspected periodically, as well as any time there are output anomalies reported by the patient.
Plug the distal (patient end) of the lead wire into the test connectors while observing polarity (red / black).
Plug the device end of the lead wire into the receptacle labeled LEAD WIRE TESTER.
When the POWER is on, green LED will turn on if the lead wire passes the test. LED will be red if the
lead wires are faulty and need to be replaced.
Figure 7
Service Center
Charge light
For reference only:
Charge light solid - charging in
progress.
Charge light blinks slowly -
charging completed.
Charge light flickers (blinks
rapidly) - batteries not installed
or are not charging properly.
Patient-end of lead
wire, Red-coded
Patient-end of lead
wire, Black-coded
Device-end of
lead wire
LED lights green when lead
wire is functional, and lights
red when lead wire needs
to be replaced.