When using the Neuroprobe
®
Output Tester, increase unit output intensity to 40 mA or higher in order to
significantly illuminate the LED. Also, view the LED located at the end of the output tester directly. If viewed at an
angle, the LED may be too dim to notice that it is illuminated.
INSPECT POWER SUPPLY OPERATION
1.
Connect the power cord to the power supply and to a power outlet.
2.
Connect the power supply cable to the stimulator unit (Figure 4).
Figure 4
3.
Press the power supply switch ON (if applicable, Figure 5a).
NOTE:
Newer power supply model does not have a ON/OFF switch (Figure 5b).
Figure 5a
Figure 5b
4.
Verify the CHARGE light is on or blinking (Figure 6).
5.
Power on the stimulator and verify that you hear a click of the safety circuit relay energizing.